Description
GAS201L Sulphides Gastec Detector Tubes
1. The measured out mud and soil is put in the generation tube and 5mL of distilled water are added.
In case of muddy water, it suffices to add a small amount of distilled water (measured out in a graduated cylinder) to the gas generation tube.
Put the cap on the generation tube.
4. Pull the handle of the gas sampling pump to draw in a sample of the generated hydrogen sulphide.
(By repeating this process several times, all sulphides become hydrogen sulphide several times repeatedly).
The scale in the detector tube is read, and the concentration of the sulphide in the sample is calculated.
Measuring Range | 0.002 to 0.02 mg |
Sampling Time | 45 seconds per pump stroke |
Detecting Limit | 0.0002 mg (n=1) |
Colour Change | White Brown |
Reaction Principle | Gus generation tube: Sulphides + H2SO4 H2S
Detector tube: H2S + Pb(CH3COO)2 PbS |
Coefficient of Variation | 5% (for 0.002 to 0.02mg) |
Shelf Life | 3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences |
Substance | Concentration | Interference | Change colour by itself |
Chloride ion | – | No error | No discoloration |
Chromate ion | – | No error | No discoloration |
Carbonate ion | – | No error | No discoloration |
Calibration gas generation | Static gas dilution method |
Additional Information
Weight | 0.1 kg |
---|---|
Gas Range | |
Brand | |
Configuration |
Product Downloads
Talk with the Experts
Contact us - Our team of application specialists can assist you with system design, installation, maintenance, and repair. As a systems integrator, we provide end-to-end support to ensure your facility has a reliable and effective gas detection system in place.
Sign up for more information - For more information on protecting your staff from the dangers of gas leaks, sign up for our emails.